ETHOS 5020 SPECIFICATION


[PDF]ETHOS 5020 SPECIFICATIONhttps://04646a9cf351cc0d3888-b8b406d15fe93f790abb5bf0e9ab7ab3.ssl.cf3.rackcd...

1 downloads 135 Views 78KB Size

ETHOS 5020 SPECIFICATION General Specification Display Range control Polarity Zero adjustment Over range indication Low battery indication Data hold Max hold Auto Power Off

Safety standards Operating environment Storage environment Fuse Power supply Test lead probe Dimension Weight

VERSION 1 08/17

LCD with a max. reading of 1999. Display sign of the test lead connect to socket Auto range & Manual range control Automatic negative polarity indication Automatic The “OL” or “-OL” display Display “

“ sign

Display “DH” sign Display “MH” sign When measurement exceeds 15 minutes without switching mode and pressing key, the meter will switch to standby mode. Press any key to exit standby mode. When restart the system, press and hold M.H key to disable auto power off EMC/LVD. CATIII 600V. The meter is up to the standards of IEC1010 Double Insulation, Pollution Degree 2, Overvoltage Category III Temperature 32° to104°F (0°C to 40°C), Humidity ≤ 80% RH. Temperature -4° to140°F (-20°C to 60°C), Humidity ≤ 90% RH. F0.5AL 250V 5 x 20mm, F10AL 250V 5 x 20mm 9V battery ( 6F22 ). RE10, CAIII 600V / CATIII 1000V, 10A, L=90cm 177(H) x 92(W) x 40(D) mm Approx. 320g (including battery)

Distributed by MTi Email: [email protected] Web: www.ethos-instruments.co.uk

Electrical Specification Accuracies are ± (% of reading + number in last digit) At 23 ± 5°C, ≤75% RH.

2.2.1 DC Voltage Range 200mV 2V 20V 200V 600V

Accuracy ± (0.5%+2)

± (0.8%+2)

Resolution 0.1mV 1mV 10mV 100mV 1V

Overload protection: 600V DC or AC rms Impedance: 10MΩ, More than 100MΩ on 200mV range

2.2.2 AC Voltage Range 2V 20V 200V 600V

Accuracy ± (0.8%+3)

± (1.2%+3)

Average sensing, calibrated to rms of sine wave Frequency: 40~500Hz Overload protection: 600V DC or AC rms Impedance: 10MΩ VERSION 1 08/17

Resolution 1mV 10mV 100mV 1V

Distributed by MTi Email: [email protected] Web: www.ethos-instruments.co.uk

2.2.3 DC Current Range 200μA 2000μA 20mA 200mA 2A 10A

Accuracy ± (1.2%+2)

± (2.0%+3)

Resolution 0.1μA 1μA 10μA 100μA 1mA 10mA

Overload protection: 0.5A/250V, 10A/250V fuse Continuous testing the maximum steady-state current of 1A

2.2.4 AC Current Range 200μA 2000μA 20mA 200mA 2A 10A

Accuracy ± (1.5%+3)

± (2.5%+5)

Resolution 0.1μA 1μA 10μA 100μA 1mA 10mA

Average sensing, calibrated to rms of sine wave Frequency: 40~500Hz Overload protection: 0.5A/250V, 10A/250V fuse Continuous testing the maximum steady-state current of 1A

2.2.5 Resistance Range 200Ω 2kΩ 20kΩ 200kΩ 2MΩ 20MΩ

Accuracy ± (1.0%+1Ω) ± (1.0%+2)

± (1.5%+3)

Resolution 0.1Ω 1Ω 10Ω 100Ω 1kΩ 10kΩ

Overload protection: 250V DC or AC rms

VERSION 1 08/17

Distributed by MTi Email: [email protected] Web: www.ethos-instruments.co.uk

2.2.6 Diode and Audible continuity test Range

Description Display read approximately forward voltage of diode Built-in buzzer sounds if resistance is less than 30Ω

Test condition Forward DC current approx. 1mA Reversed DC voltage approx. 1.5V Open circuit voltage approx. 0.5V

Overload protection: 250V DC or AC rms

2.2.7 Battery test Range 1.5V 9V

Accuracy ±(5.0%+5 )

Load current 100mA 5mA

Resolution 1mV 10mV

Overload protection: 0.5A/250V fuse

VERSION 1 08/17

Distributed by MTi Email: [email protected] Web: www.ethos-instruments.co.uk