General Specifications - Yokogawa Electric Corporation


Jul 9, 2012 - ...

3 downloads 197 Views 950KB Size

General Specifications

Differential pH Sensor Model PH18

GS 12B6J4-01E-A The Model PH18 Differential pH Sensor is unique and offers the possibility of maintenance-free operation for the correct application. Construction of the enamel coated probe is based on a rigid steel rod. A blue base enamel is overlaid with two yellow bands of sensitive enamel. One pH sensitive enamel and the second a sodium ion enamel. They are combined with an integral temperature sensor, to form a probe that does not have any internal solutions. The differential measuring principle combines the normal potential generated by the pH enamel with the potential from the sodium membrane. In suitable applications the level of sodium ions creates a stable reference voltage. Hence the measurement can be made without a conventional reference electrode, and drawbacks associated with a liquid junction. The sensitive enamels are directly bonded to a metallic substrate, eliminates the need for conventional internal reference elements. This unique construction makes the PH18 independent of the drift typically caused by conventional reference elements as they age or become contaminated. When used in a system which is cleaned and/or sterilized with hot water or steam, the enamels produce a signal which has long term stability unrivalled by conventional systems.

APPLICATION Process where KCl is forbidden Process containing a low water content High process temperature and pressure Process with rapid pressure/temperature changes Fermentation processes Dairy product manufacture

FEATURES Sterilizable All enamel construction No reference fill solution or junction Easy installation and commissioning No routine maintenance Flexible installation options 140 ºC, 15 bar (284 ºF, 214 psi)

SYSTEM CONFIGURATION

Sensors Yokogawa Corporation of America 2 Dart Road, Newnan, Georgia U.S.A. 30265 Tel: 1-800-258-2552 Fax: 1-770-254-0928

GS 12B6J4-01E-A ©Copyright April 2003 GS 12B6J4-01E-A 4th Edition July 9, 2012

2nd-Edition

 2

2

INTRODUCTION INTRODUCTION

The PH18 electrode.It Itrelies relies The PH18isisnot notaaconventional conventional electrode. on on twotwo dissimilar membranes generate signal. dissimilarion ionsensitive sensitive enamel enamel membranes to to generate the the signal. The measuring likeaapH pHelectrode, electrode, and reference The measuringelement element is is like and thethe reference likelike a sodium a sodiumion ionelectrode. electrode.

Measuring Enamel mV vs. pH in solution containing 0.1N Na+ Alkaline error

Therefore be used usedinincombination combination with Thereforethe thePH18 PH18can can only only be with pH pH meters that meters thatfeature: feature: 1 Dual highhigh impedance inputs 1 Dual impedance inputs 2 Adjustable setting forfor Isothermal 2 Adjustable setting IsothermalPoint Point 3 Adjustable temperature coefficient 3 Adjustable temperature coefficient

Operating range

The Yokogawa and PH402/450 The Yokogawamodels modelsPH202/FLXA21 PH202 and PH450 satisfy these satisfy these requirements. requirements. In both cases arebonded bonded directly to the In both casesthe themembrane membrane enamels enamels are directly to the metal substrate solution. metal substratewith withno nofilling filling solution.

pH of sample

Fig. 1

Acid error

The response curves Fig. 1 (pH), Fig. 2 (reference) and Fig. 3 The response curves Fig. 1 (pH), Fig. 2 (reference) and Fig. 3 (application range) show the way in which the potential is (application range) show the way in which the potential is generated at each element and combined in the output of the generated at each element and combined in the output of the sensor.

Reference Enamel mV vs pH in solution containing 0.1N Na+

sensor. Fig. 1 above pH11 the effect of sodium on the pH membrane Fig. 1 above pH11 the effect of sodium on the pH membrane starts to cause non-linearity (sodium error) starts to cause non-linearity (sodium error) Fig. 2 below pH3 the high hydrogen ion content causes a Fig. 2 below pH3 the high hydrogen ion content causes a change in the reference response change in the reference response a “linear” response to pH is obtained between 3 and 11 Fig.Fig. 3 3a “linear” response to pH is obtained between 3 and 11 pH pH a plot of referencevoltage voltageagainst against sodium Fig.Fig. 4 4a plot of reference sodiumcontent content The revolutionarymeasuring measuring principle some bigbig advantages. The revolutionary principlehas has some advantages. The absence of filling solutions and reference junctions virtually The absence of filling solutions and reference junctions virtually eliminatesthe theproblems problems caused caused by of the eliminates byaging agingand andpollution pollution of the referencesensor. sensor.Regular Regular cleaning virtually reference cleaningofofthe thesensor sensor virtually eliminatesdrift, drift,and andthe the sensor sensor benefits long working eliminates benefitsfrom froma very a very long working life. It is vital, however, to fit the sensor to the application life. It is vital, however, to fit the sensor to the application correctly. Thespecial specialnature nature of dictates correctly. The of the the reference referenceelement element dictates there mustbe beaacertain certain sodium sodium level Fig.Fig. 4 4 thatthat there must levelininthe theprocess. process. shows a plot reference voltage voltage against content. shows a plot ofofreference againstsodium sodium content. Because theexponential exponential nature it isit plain thatthat Because ofofthe natureofofthe theresponse, response, is plain solution, the the reference above about0.5 0.5NNNa+ Na (30g above about (30g NaCl/l) NaCl/l)ininthethe solution, reference output tends towardsaaconstant constant level. level. Of thethe output tends towards Ofcourse coursewhen when sodium concentrationremains remains constant thethe sodium concentration constantinina aprocess process referencevoltage voltagewill will also also be lower levels of of reference be constant constantatatmuch much lower levels sodium. sodium.

Operating range

pH of sample

Fig. 2 PH18 system output pH reading vs. pH in solution containing 0.1N Na+

+

It because is becauseofofthe theneed need to to evaluate evaluate the of of thethe It is thechemistry chemistry process, thatititisisnecessary necessary to Data process, that tohave havean anApplication Application Data Sheet (page7 7ofofthis thisGS) GS) completed completed before forfor thisthis Sheet (page beforeapproval approval sensor canbe bemade. made. sensor can

Operating range

pH of sample

Fig. 3

The mechanicalconstruction construction of means thatthat it it The mechanical of the thesensor sensoralso also means may usedininprocesses processes involving involving both andand may bebe used bothhigh hightemperatures temperatures pressures.By Byeliminating eliminating the thethe sensor is truly pressures. the filling fillingsolutions, solutions, sensor is truly robust and can severethermal thermal shocks would robust and caneven evenwithstand withstand severe shocks thatthat would ruinruin most systems. most systems.

Reference Enamel mV vs Na+ content

The stainless adapterforms formsthethe liquid earth The stainlesssteel steel mounting mounting adapter liquid earth (solution neededtotoensure ensure best stability (solutionground) ground)connection connection needed best stability of of measurement. thisconnection connection diagnostic measurement.EXA EXA also also uses uses this in in thethe diagnostic circuit. circuit. The PH18 isisa adifferential sensor.ItItdoes doesnot not measure The PH18 differential pH pH sensor. measure absolute pH applications.It It does, however, absolute pHexcept exceptin inlimited limited applications. does, however, measure a asingle pointaccurately, accurately, repeatably measure singlecontrol control point repeatably andand with minimum with minimummaintenance. maintenance. Fig. 4

All Rights Reserved. Copyright © 2003, Yokogawa Corporation of America. Subject to change without notice.

GS 12B6J4-01E-A

GS 12B6J4-01E-A

July 9, 2012

GENERAL GENERAL SPECIFICATIONS SPECIFICATIONS

GENERAL GENERAL SPECIFICATIONS SPECIFICATIONS

 3

Model Model PH18 PH18 Differential Differential pH pH Sensor Sensor Temperature :: Pt1000 Temperature sensor sensor Pt1000 Ω Ω RTD RTD Wetted parts : pH sensor enamel Wetted parts : pH sensor -pH -pH sensitive sensitive enamel + :: Reference GENERAL SPECIFICATIONS sensitive Reference sensor sensor -- Na Na+ sensitive enamel enamel Liquid earth Model PH18 Differential:: pH Sensor Liquid earth via via SS SS adapter adapter : O-rings (EPDM) Temperature sensor : O-rings : Pt1000 RTD (EPDM) :: Adapters (SS Wetted parts : pH sensor -pH sensitive enamel Adapters (SS 316) 316) Max. :: 3 11 actual be Reference Na+ will sensitive Max. measuring measuring range range 3:to to 11 pH pH (The (Thesensor actual -range range will be advised with enamel advised with reference reference to to the the completed application sheet) : Liquid earth via SSdata adapter completed application data sheet) Temperature :: 0 140 (284 : Temperature range range 0 to toO-rings 140 ºC ºC(EPDM) (284 ºF) ºF) Pressure range : -1 to 15 Bar (214 psi) Pressure range : -1 to 15 Bar (214 psi)

Model Model WU18 WU18 cable cable for for model model PH18 PH18 Differential Differential pH pH Sensor Sensor Max. temperature : 110 °C (230 °F) Max. temperature : 110 °C (230 °F) Material :: Thermoplastic Material Thermoplastic Rubber Rubber (T.P.R.) (T.P.R.) GENERAL SPECIFICATIONS Color : Blue Color : Blue Shipping details sensor and Shipping details sensor and adapter adapter Model WU18 cable for model PH18 pH Sensor Package : wxhxd 350 xxDifferential 220 xx 110 Package : wxhxd 220°F) 110 mm mm Max. temperature : 110 350 °C (230 Weight : approx. 1.4 kg Weight : approx. 1.4 kg Material : Thermoplastic Rubber (T.P.R.) Shipping details cable Shipping details cable Color : Blue350 x 220 x 110 mm Package : wxhxd Packagedetails sensor and : wxhxd 350 x 220 x 110 mm Shipping adapter Weight :: 2 approx. 0.6 kg 2: m. m. approx. 0.6 kg x 110 mm Weight Package wxhxd 3500.9 x 220 :: 5 m. approx. kg 5 m. approx. 0.9 kg Weight : approx. 1.4 kg :: 10 10 m. m. approx approx 1.4 1.4 kg kg

Sheet. Any and all information received by Yokogawa will be treated in treated in the the strictest strictest confidence. confidence. To To maintain maintain traceability, traceability, the the NOTE: completed Application Data Sheet will form part of of completed Application Data Sheet will form part of the the contract contractlocal of The use of this sensor is highly application specific.Your sale. sale.

MODEL AND SUFFIX CODE

: Adapters (SS 316) Max. measuring range : 3 to 11 pH (The actual range will be NOTE: NOTE: advised withspecific.Your reference to the The application local The use use of of this this sensor sensor is is highly highlycompleted application application specific.Yourdata localsheet) Yokogawa office to advise on the suitability Yokogawa sales sales office will will be be: pleased pleased toºC advise on Temperature range 0ofto 140 (284Application ºF)the suitability of your application, on receipt the completed Data of your application, on receipt of the completed Application Pressure to 15 by BarYokogawa (214 psi)will beData Sheet. Anyrange and all information: -1 received

Yokogawa sales office will be pleased to advise on the suitability ofYokogawa your application, on receipt of the completed Application Data Yokogawa offers offers no no function function guarantee guarantee for for applications applications where where Sheet. Any and allsheet information7) received by Yokogawa will be the the attached attached data data sheet (page (page 7) has has not not been been satisfactorily satisfactorily treated in the strictest confidence. To maintain traceability, the completed. completed. This This does does not not affect affect the the normal normal Yokogawa Yokogawa warranty warranty covering in workmanship. completed Application Dataor Sheet will form part of the contract of covering defects defects in materials materials or workmanship. sale. Yokogawa offers no function guarantee for applications where the attached data SUFFIX sheet (page 7) has not been satisfactorily MODEL CODE MODEL AND AND SUFFIX CODE completed. This does not affect the normal Yokogawa warranty covering defects materials or workmanship. Model Suffixin Option Description Model Suffix Option Description Code Code

Shipping details cable Package : wxhxd 350 x 220 x 110 mm Weight : 2 m. approx. 0.6 kg MODEL 5 m. approx. 0.9 kg MODEL AND AND SUFFIX SUFFIX: CODE CODE : 10 m. approx 1.4 kg

Model Model

Model

WU18 WU18

WU18

Suffix Suffix

Suffix Code Code Code -03 -03 -05 -05 -03 -10 -10

Option Option

Option Code Code Code

-05 -10

material

Instruction manual

-SA

-SA Suffix Code -E -E

Code Code

Option Code

-E

Compatible with 25mm Compatible with 25mm Description process connection process connection Ethylene-propylene (EPDM) Ethylene-propylene (EPDM)

MEASURE MEASURE SCREEN SCREEN SCREEN SCREEN

Model Ph18 Differential pH Sensor

-E -SA -E

-E

03 connection cable cable 03 meter meter connection 05 meter meter connection connection cable cable 05 03 connectioncable 10 meter meter connection connection 10 meter cable

05 meter connection cable 10 meter connection cable

odel Ph18 Differential pH Sensor MODEL AND SUFFIXM MCODE odel Ph18 Differential pH Sensor PH18 PH18 Sensor Sensor Model mounting mounting O-ring O-ring PH18 material material Instruction Sensor Instruction manual mounting manual Options O-ring Options

Description Description

Description

REFERENCE REFERENCE

English language Compatible with 25mm process English language

K1520EK K1520EK K1520EJ K1520EJ

K1525YA K1525YA 237230 237230

connection

Angled weld-in adapter(EPDM) (SS316) Ethylene-propylene Angled weld-in adapter (SS316) Staight weld-in adapter (SS316) Staight weld-in adapter (SS316) Adapter 1.25” NPT (SS316) Adapter NPT (SS316) English1.25” language Blind plug SS316 (EPDM O-ring) Blind plug SS316 (EPDM O-ring)

Options

K1520EK Angled weld-in adapter (SS316) K1520EJ Staight weld-in adapter (SS316) Note: Note: The sensor is supplied with cable connector. Adapter For first installation must be M1289BA 1” NPTcable (SS316) The sensor is supplied with cable connector. For first installation cable must be specified as well. 237230 Blind plug SS316 (EPDM O-ring) specified as well.

LIQUID EARTH LIQUID EARTH TEMPERATURE TEMPERATURE

Model PH18 Model PH18 sensor sensor

TEMPERATURE TEMPERATURE

Fig. Fig. 5 5 Connection Connection diagram diagram WU18 WU18 for for PH402 PH402 and and PH202 PH202

Fig. 5 Connection diagram WU18 for PH202/FLXA21 and PH402/PH450

Note: Note: The material certificate 3.1.B is supplied with the sensor and the options The material certificate 3.1.B is supplied with the sensor and the options

Measuring Measuring Range Range

Degree DegreeºCºC

Note: The sensor is supplied with cable connector. For first installation cable must be specified as well. Note: The material certificate 3.1.B is supplied with the sensor and the options

1 1 2 2

pH of sample pH of sample

Fig. Fig. 6. 6. Corrosion Corrosion curve curve of of the the Model Model PH18 PH18 Enamel Enamel

GS GS 12B6J4-01E-A 12B6J4-01E-A All Rights Reserved. Copyright © 2003, Yokogawa Corporation of America. Subject to change without notice.

GS 12B6J4-01E-A

July 9, 2012

 4

4

DIMENSIONS PH18-SA in mm (inches)

PH18SA

237230

K1520EJ 237574

K1520EK 237202

M1289BA

Fig. 8. External dimensions of the PH18-SA

Fig. 7. PH18-SA and the adapters

55 (2.17)

51 (2.09)

44 (1.73)

22.5 (0.89)

26 (1.02)

1”

242574 K1520EJ

237202 K1520EK

M1289BA M1289BA

237230 237230

Fig. 9. Options of the PH18-SA All Rights Reserved. Copyright © 2003, Yokogawa Corporation of America. Subject to change without notice.

GS 12B6J4-01E-A

GS 12B6J4-01E-A

July 9, 2012

 5 5

PH18 INSTALLATION EXAMPLES

MODEL PH 18 ELEMENTS

Socket for connection cable

Primary adapter

pH enamel

Reference enamel

Fig. 13. Example A. Straight weld-in adapter through a vessel wall. Example B. 1” Screw-in adapter with existing pipe nipple through a vessel wall. Example C. Angled weld-in adapter in large bore pipe. Example D. Angled weld-in adapter mounted in a bend Example E. Screw or weld-in adapter mounted in a large bore pipe. Example F. Screw or weld-in adapter mounted on top of a large bore pipe.

Fig. 14. The primary adapter is used to make a connection to the process used by the transmitter as the liquid earth (solution ground). This is needed for optimum stability of the measurement and is also used in the impedance checking circuit.

Note 1: When measuring in plastic tanks or pipes, ensure that the adapter is wetted by the process. Avoid installations where an air pocket can be created (see fig 12 F). This isolates the adapter, and hence loses the liquid earth connection. Note 2: Flow rate from the side of sensor (fig. 12 (C & D) ) should not exceed 2 metres/second in low viscosity fluids. In high viscosity fluids (>5cP) use only installation as shown in fig. 12 (D).

Fig. 15. Should be used to show process variations. The changes in pH and sodium content should be plotted against time for batch processes. Changes in temperature and pressure can also be plotted if appropriate. Where CIP is used, plot the cleaning cycle, and any sterilization stages. This leads to a clearer understanding of the process.

All Rights Reserved. Copyright © 2003, Yokogawa Corporation of America. Subject to change without notice.

GS 12B6J4-01E-A

July 9, 2012 GS 12B6J4-01E-A

 6

6 6

SPARE PARTS SPARE PARTS PH18-SA Caution - Inappropriate handling can cause damage. Striking or scratching the sensor against hard surfaces such as Caution - Inappropriate handling cancan cause damage. Caution cause steel, stone,- Inappropriate glass or ceramic,handling may cause damage todamage. the enamel. Striking ororscratching thethe sensor against hardhard surfaces such such as as Striking scratching sensor against surfaces Such damage may not affect the sensor performance steel, stone, glass oror ceramic, may cause damage to the enamel. steel, stone, glass ceramic, may cause damage to the enamel. immediately, but after prolonged exposure to the process, flaking Such damage may not affect thethe sensor performance damage may not affect sensor performance ofSuch the enamel may occur. immediately, prolonged exposure to the process, flaking immediately,but butafter after prolonged exposure to the process, flaking of the enamel may occur.

PH18-SA

of the enamel may occur.

6

SPARE PARTS PH18-SA

Note - Enable impedance checks in the PH202 or PH450 Caution - Inappropriate handling damage. Note - Enable impedance checks incause the PH202/FLXA21 and The sensor impedance checking cancan give early warning of Note - Enable impedance checks in the PH202 or PH450 Striking or scratching the sensor against hard surfaces PH420/PH450 damage to the enamel layers. If the transmitter signals ansuch as The impedance checking can givegive earlyearly warning steel, stone, glass ceramic, may cause damage theofenamel. Thesensor sensor impedance checking can of impedance failure oforthe PH18, the sensor should betowarning removed damage to the enamel layers. If the transmitter signals an an Such damage may not affect the sensor performance damage to the enamel If the(within transmitter signals from the process, as soon layers. as possible 24 hours). The loss impedance failure of of the PH18, the sensor should be removed but after exposure to avoided. the process, flaking impedance failure the PH18, the sensor should be removed ofimmediately, significant amounts ofprolonged enamel may thus be from the process, as soon as possible (within 24 hours). The loss of the enamel may occur. from the process, as soon as possible (within 24 hours). The loss of significant amounts of enamel may thus be avoided.

Size of the O-Ring A= 17.0 x 2.5 B= 20.0of the x 2.5O-Ring Size C= A=25.0 17.0 x x1.5 2.5 D= 17.0 B= 20.0 x x3.0 2.5 C= 25.0 x 1.5 D= 17.0 x 3.0

of significant amounts of enamel may thus be avoided.

ACCESSORIES Spare parts for model PH18 differential pH sensor ACCESSORIES

Materials EPDM-FDA- approved EPDM-FDA-approved Materials Viton A EPDM-FDAapproved EPDM EPDM-FDA-approved Viton A EPDM

Fig. 16. Location of the O-rings

Fig. 16. Location of the O-rings Spare parts for model PH18 differential pH sensor Number Description Quantity Note - Enable impedance checks in the PH202 or PH450 The sensor impedance checking can give early warning of O-ring set contains Qty 1 each of A, B, C&D from figure 16 M1263XY Number Description Quantity damage to the enamel layers. If the transmitter signals an 242574 Straight weld-in adapter 1 O-ring setPH18, contains 1 each of A, C&D from figure 16 M1263XYfailure impedance of the the Qty sensor should beB, removed of the O-Ring Materials 237202 Angled weld-in adapter 11Size from the process, as soon as possible (within 24 hours). The loss 242574 Straight weld-in adapter A= 17.0 x 2.5 EPDM-FDA- approved B= 20.0 x 2.5 EPDM-FDA-approved K1525YA AAngled dapter of 1.25 " NPT 11C= 25.0 x 1.5 of significant amounts enamel may thus be avoided. 237202 weld-in adapter Viton A D= 17.0 x 3.0 EPDM Blind plug and mounting out 237230 1 K1525YA Adapter 1.25 " NPT 1 ACCESSORIES Blind plug and mounting outpH sensor 237230 1 Spare parts for model PH18 differential Number

Description

M1263XY

O-ring set contains Qty 1 each of A, B, C&D from figure 16

ACCESSORIES K1520EJ

Straight weld-in adapter

SpareK1520EK parts for model PH18 differential pH sensor Angled weld-in adapter M1289BADescription Adapter 1” NPT Number 237230 M1263XY 242574 237202 K1525YA

237230

Quantity 1 Fig. 16. Location of the O-rings

Blind plug and outA, B, C&D from figure 16 O-ring set contains Qtymounting 1 each of Straight weld-in adapter Angled weld-in adapter Adapter 1.25 " NPT Blind plug and mounting out

1

Quantity

1 1

1 1 1 1

All Rights Reserved. Copyright © 2003, Yokogawa Corporation of America. Subject to change without notice.

GS 12B6J4-01E-A

July 9, 2012

SC24V or PH18 Application Data Sheet Customer:

Please complete a seperate form for each process

Tag No:

stream to be analyzed and return to the attention of:

Address:

Yokogawa Corporation of America Analytical Business Unit

Contact:

2 Dart Road

Telephone:

Newnan,

Email:

Georgia

30265 - 1040

FAX - (770) 304-1613

Authorization No.

Rep Name: 1. PROCESS DATA: Sensor Selection:

PH18

SC24V

Application: Operation

Chemical Batch

Power Control

Pharmaceutical Continous

Pulp & Paper Monitor

Other Other

Type of Solution:

2. PROCESS CONDITIONS: Descrition:

pH Value Conductivity Value Sodium Concentration Temperature (°C/°F) Pressure (psig) Flow Rate (ft/sec)

Min Min Min Min Min Min Min Min

Max Max Max Max Max Max Max Max

Norm Norm Norm Norm Norm Norm Norm Norm

Concentration of Solids: Type(s):

Concentration of Organics: Type(s):

Problems: Describe:

Fouling

Abrasion

Coating

Poisoning

Other

Cleaning/Sterilization: Describe:

w/ Product

Caustic

Steam

Hot Water

Other

Retractable On Line tank

Immersion Sample Line

Other Flanged

Kynar

Kalrez

3. INSTALLATION DATA: Mounting:

Insertion Off Line

Wetted Materials:

316SS

Flow Through On Line Pipe EPDM

ft

Distance to Converter: Instrument:

General Purpose

Power Supply:

115 VAC

Intrinsic Safe

24 VDC

All Rights Reserved. Copyright © 2003, Yokogawa Corporation of America. Subject to change without notice.

GS 12B6J4-01E-A

July 9, 2012

7

SPARE PARTS PH18-SA Caution - Inappropriate handling can cause damage. Striking or scratching the sensor against hard surfaces such as steel, stone, glass or ceramic, may cause damage to the enamel. Such damage may not affect the sensor performance immediately, but after prolonged exposure to the process, flaking of the enamel may occur.

Note - Enable impedance checks in the PH202 or PH402 The sensor impedance checking can give early warning of damage to the enamel layers. If the transmitter signals an impedance failure of the PH18, the sensor should be removed from the process, as soon as possible (within 24 hours). The loss of significant amounts of enamel may thus be avoided.

Size of the O-Ring A= 17.12 x 2.62 B= 20.30 x 2.62 C= 25.12 x 1.78 D= 17.12 x 2.62

ACCESSORIES

Fig. 16. Location of the O-rings

Spare parts for model PH18 differential pH sensor

Number

Description

K1500BJ K1500BR K1500BS K1500BT K1522BR K1522EQ K1522ET K1522ES K1522MV

Set O-rings 20.3 X 2.62 (EPDM) unit Set O-rings 17.12 X 2.62 (EPDM) unit Set O-rings 25.12 X 1.78 (EPDM) unit Nut /SWR Straight weld-in adapter /SWA Angled weld-in adapter /SBS Adapter ISO 7/1 - R1, JIS 1" /SNS Adapter 1" NPT /BSA1 Blind plug

EPDM K1500BR K1500BJ K1500BS K1500BR

Quantity 5 5 1 1 1 1 1 1 1

YOKOGAWA ELECTRIC CORPORATION World Headquarters 9-32, Nakacho, Musashino-shi, Tokyo 180-8750 JAPAN Tel: 81-422-52-5555 Fax: 81-422-55-0461 www.yokogawa.co.jp

Represented by:

YOKOGAWA CORPORATION OF AMERICA 2 Dart Road, Newnan, GA 30265 USA Tel: 800-258-2552 Fax: 1-770-254-0928 www.us.yokogawa.com YOKOGAWA EUROPE B.V. Databankweg 20 3821 AL Amersfoort THE NETHERLANDS Tel: 31-33-4641 611 Fax: 31-33-4641 610 www.yokogawa.com/eu YOKOGAWA ELECTRIC ASIA PTE, LTD. Bedok Factory 5 Bedok South Road SINGAPORE 469270 Tel: 65-241-9933 Fax: 65-241-9537 www.yokogawa.com.sg

GS 12B6J4-01E-A GS GS 12B6J4-01E-A 12B6J4-E-H Subject change without notice Subject tosubject change withoutwithout noticenotice Contentsto to change Copyright 2003 Copyright ©2003, 2003YCA Copyright©© Printed in USA All All Rights Rights Reserved. Reserved. Copyright Copyright © © 2003, 2003, Yokogawa Yokogawa Corporation Corporation of of America. America. Subject Subject to to change change without without notice. notice.

PrintedininThe TheUSA, USA,04-1207 01-705 (A) Printed (A) II

GS 12B6J4-01E-A GS GS 12B6J4-01E-A 12B6J4-01E-A

August July 4, 9, 2011 2012