Range: 1W to 200 W Sensors for Laser Power and


[PDF]Range: 1W to 200 W Sensors for Laser Power and...

0 downloads 89 Views 1MB Size

Sensors for Laser Power and Position Sensing ABSORBERS

Range: 1W to 200 W Features: · High Beam Position Accuracy · Highest Power Density on SHC Coating · HPB coating is suitable for Excimer Lasers Model

QA-200-D40-HPB

QA-200-D40-SHC

Max. Average Power

200 W

200 W

Max. Intermittent Power (1)

250 W

250 W

1W

1W

Power Resolution

10 mW

10 mW

Noise Equivalent Power (NEP)

Min. Power

10 mW

10 mW

Response Time

2 sec

2 sec

Power Calibration Uncertainty

± 3%

± 3%

Power Linearity (2)

± 1%

± 1%

Beam Position Accuracy

100µm

100µm

Beam Position Resolution

100µm

100µm

1W

1W

40 mm

40 mm

SENSORS

Power Mode

Minimum Power (3)

MONITORS & SW

Beam Sensing Mode

Aperture

HPB

SHC

0.19 - 11 µm

0.19 - 11 µm

0.19 - 2.1 µm, 2.94µm, 9 - 11 µm

0.25 - 1.1 µm, 9 - 11 µm

Type Absorber Spectral Range Calibration Spectral Range Max Power Density (4) Max Energy Density

(4)

7 kW/cm² @200 W

28 kW/cm² @200 W

5ms pulse width: 36 J/cm² 10µs pulse width: 1.2 J/cm² 10ns pulse width: 0.3 J/cm²

5ms pulse width: 115 J/cm² 10µs pulse width: 4 J/cm² 10ns pulse width: 1 J/cm²

Forced Air with Fan (a)

Forced Air with Fan (a)

USB/RS232 SENSORS

Absorber Specs

Cooling Weight Dimension

1.2 kg

1.2 kg

100 x 100 x 85 mm

100 x 100 x 85 mm

1.5 m - DB25

1.5 m - DB25

Light Duty Stand Included

Light Duty Stand Included

a). 12V DC Power Supply Included

a). 12V DC Power Supply Included

Cable lenght - connector Stand and Post

OEM SOLUTIONS

General Characteristics

Notes

POWER PROBES

(1). 2 minutes max (2). Detector centrally irradiated @50% of useful surface. (3). necessary for beam position sensing. (4). Damage thresholds also depend on power level. Please see damage graphs

QA-200-D40-HPB

QA-200-D40-SHC

Light Duty Stand

Via Burona, 51 - Vimodrone (Milano) - ITALY Ph: +39 02 27 40 02 36 Fax: +39 02 25 02 91 61 For latest updates please visit our website: www.laserpoint.eu

35

Sensors for Laser Power and Position Sensing ABSORBERS

Range: 10W to 3kW Features: · High Beam Position Accuracy · Water cooled heads to 3000 W · Corrosion-Proof Water-Cooled Sensors Model

QW-1500-D40-HPB

QW-1500-D40-SHC

QW-3000-D55-HPB

QW-3000-D55-SHC

Max. Average Power

1500 W

1500 W

3000 W

3000 W

Max. Intermittent Power (1)

2250 W

2250 W

4500 W

4500 W

10 W

10 W

20 W

20 W

Power Resolution

100 mW

100 mW

100 mW

100 mW

Noise Equivalent Power (NEP)

Power Mode

SENSORS

Min. Power

100 mW

200 mW

250 mW

250 mW

Response Time

4 sec

4 sec

5 sec

5 sec

Power Calibration Uncertainty

± 5%

± 5%

± 5%

± 5%

± 1.5%

± 1.5%

± 2%

± 2%

Beam Position Accuracy

100µm

200µm

300µm

300µm

Beam Position Resolution

100µm

100µm

100µm

100µm

Minimum Power (3)

7.5 W

7.5 W

15 W

15 W

40 mm

40 mm

55 mm

55 mm

HPB

SHC

HPB

SHC

0.19 - 11 µm

0.19 - 11 µm

0.19 - 11 µm

0.19 - 11 µm

0.19 - 2.1 µm, 2.94µm, 9 - 11 µm

0.25 - 1.1 µm, 9 - 11 µm

0.19 - 2.1 µm, 2.94µm, 9 - 11 µm

0.25 - 1.1 µm, 9 - 11 µm

2.4 kW/cm² @1 kW

7 kW/cm² @1 kW

1.8 kW/cm² @2kW

3.6 kW/cm² @2kW

5ms pulse width: 36 J/cm² 10µs pulse width: 1.2 J/cm² 10ns pulse width: 0.3 J/cm²

5ms pulse width: 115 J/cm² 10µs pulse width: 4 J/cm² 10ns pulse width: 1 J/cm²

5ms pulse width: 36 J/cm² 10µs pulse width: 1.2 J/cm² 10ns pulse width: 0.3 J/cm²

5ms pulse width: 115 J/cm² 10µs pulse width: 4 J/cm² 10ns pulse width: 1 J/cm²

Water (a)

Water (a)

Water (a)

Water (a)

Power Linearity (2)

POWER PROBES

OEM SOLUTIONS

USB/RS232 SENSORS

MONITORS & SW

Beam Sensing Mode

Absorber Specs Aperture Type Absorber Spectral Range Calibration Spectral Range Max Power Density

(4)

Max Energy Density

(4)

General Characteristics Cooling Weight Dimension Cable lenght - connector Stand and Post

1.1 kg

1.1 kg

2.3 kg

4.2 kg

Ø 116 x 44 mm

Ø 116 x 44 mm

Ø 148 x 55 mm

Ø 148 x 55 mm

5 m - DB25

5 m - DB25

5 m - DB25

5 m - DB25

Heavy Duty Stand Included

Heavy Duty Stand Included

Heavy Duty Stand Included

Heavy Duty Stand Included

Notes (1). 2 minutes max (2). Detector centrally irradiated @50% of useful surface. (a). Water 4 liter/min (@ 22°C); (a). Water 4 liter/min (@ 22°C); (a). Water 5 liter/min (@ 22°C); (a). Water 5 liter/min (@ 22°C); (3). necessary for beam position senadmissible rate of temperature admissible rate of temperature admissible rate of temperature admissible rate of temperature sing. variation < 1 °C/min variation < 1 °C/min variation < 1 °C/min variation < 1 °C/min (4). Damage thresholds also depend on power level. Please see damage graphs for more details

QW-1500-D40-HPB

QW-1500-D40-SHC

QW-3000-D55-HPB

Via Burona, 51 - Vimodrone (Milano) - ITALY Ph: +39 02 27 40 02 36 Fax: +39 02 25 02 91 61 For latest updates please visit our website: www.laserpoint.eu

QW-3000-D55-SHC

36

Sensors for Laser Power and Position Sensing ABSORBERS

Range: 30W to 6kW Features: · High Beam Position Accuracy · Water cooled heads to 3000 W · Corrosion-Proof Water-Cooled Sensors Model

QW-6000-D55-SHC

Power Mode Max. Average Power

6000 W

Max. Intermittent Power (1)

9000 W

Power Resolution

100 mW

Noise Equivalent Power (NEP)

500 mW

Response Time

3.5 sec

Power Calibration Uncertainty

± 5%

Power Linearity (2)

± 2%

SENSORS

30 W

Min. Power

Beam Position Accuracy

300µm

Beam Position Resolution

100µm

Minimum Power (3)

MONITORS & SW

Beam Sensing Mode

30 W

Absorber Specs 55 mm

Aperture

SHC

Type

0.19 - 11 µm

Absorber Spectral Range

Max Power Density

(4)

Max Energy Density

USB/RS232 SENSORS

0.25 - 1.1 µm, 9 - 11 µm

Calibration Spectral Range

4 kW/cm² @5kW 5ms pulse width: 115 J/cm² 10µs pulse width: 4 J/cm² 10ns pulse width: 1 J/cm²

(4)

General Characteristics Water (a)

Cooling

4.2 kg

Weight

OEM SOLUTIONS

Ø 148 x 55 mm

Dimension

5 m - DB25

Cable lenght - connector

Heavy Duty Stand Included

Stand and Post Notes

QW-6000-D55-SHC

(a). Water 8 liter/min (@ 22°C); admissible rate of temperature variation < 1 °C/min

POWER PROBES

(1). 2 minutes max (2). Detector centrally irradiated @50% of useful surface. (3). necessary for beam position sensing. (4). Damage thresholds also depend on power level. Please see damage graphs

Heavy Duty Stand

Via Burona, 51 - Vimodrone (Milano) - ITALY Ph: +39 02 27 40 02 36 Fax: +39 02 25 02 91 61 For latest updates please visit our website: www.laserpoint.eu

37